80K-15
Electronic Air Cleaner Probe
Instructions
The 80K-15 Electronic Air Cleaner Probe (the probe or Product) is a low
energy probe that is able to extend the voltage measuring capability of an
ac/dc voltmeter or multimeter.
Properly connected, the probe is rated to 15 kV dc or 10 kV ac rms. The
use of the probe is restricted to low-energy applications that are energy-
limited circuits such as electrostatic air cleaners.
The probe is a precision 1000:1 voltage divider formed by two matched
resistors. The high input impedance offered by these resistors minimizes
circuit loading and optimizes measurement accuracy.
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
•Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
•Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
•Canada: 1-800-36-FLUKE (1-800-363-5853)
•Europe: +31 402-675-200
•Japan: +81-3-6714-3114
•Singapore: +65-6799-5566
•China: +86-400-921-0835
•Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com
.
To register your product, visit http://register.fluke.com
.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
.

Safety Information
A Warning identifies conditions and procedures that are dangerous to the
user. A Caution identifies conditions and procedures that can cause
damage to the Product or the equipment under test.
XW Warning
To prevent possible electrical shock, fire, or personal injury:
•Read all safety information before you use the Product.
•Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
•Examine the case before you use the Product. Look for
cracks or missing plastic.
•Do not use test leads if they are damaged. Examine the test
leads for damaged insulation and measure a known voltage.
•Do not alter the Product and use only as specified, or the
protection supplied by the Product can be compromised.
•When you do a measurement, never make body contact with
the probe tip or red portion of the Product. Always hold the
Product by its black handle.
•Do not use this probe to measure high voltages on power
distribution systems.
•Connect the protective earth alligator clip lead to a known
good earth ground.
•Do not apply more than the rated voltage between the
terminals or between each terminal and earth ground.
•Have an approved technician repair the Product.
•Do not use the Product if it is altered or damaged.
•Disable the Product if it is damaged.
Symbols
Symbol Definition
X
WARNING. HAZARDOUS VOLTAGE. Risk of electric
shock.
W
WARNING - RISK OF DANGER. Consult user
documentation.
T
Double Insulated
Earth
PConforms to European Union directives.
)
Certified by CSA Group to North American safety
standards.
~
This product complies with the WEEE Directive marking
requirements. The affixed label indicates that you must
not discard this electrical/electronic product in domestic
household waste. Product Category: With reference to
the equipment types in the WEEE Directive Annex I,
this product is classed as category 9 "Monitoring and
Control Instrumentation" product. Do not dispose of this
product as unsorted municipal waste.

Operation
Use the probe as follows:
1. Inspect the probe.
2. Connect the ground lead of the probe to earth ground.
3. Connect the probe cable to a compatible voltmeter. Use a shunt resistor
if required.
4. Select the 10 MΩ voltage range (1 V reading for 1000 V probe input).
5. Hold the probe by the black handle and connect the probe tip to the
circuit.
6. Observe the voltmeter reading.
Probe tip
Probe body
Multimeter or voltmeter
Plug
GND tab
Protective earth ground lead clip
MIN MAX
RANGE
REL
Hz %
AutoHOLD
Peak MIN MAX
4½ DIGITS
1 Second
°
C/
°
F
OFF
mA
A
mV
V
V
A
A
mA
COM
V
400mA
FUSED
10A MAX
FUSED
A
TRUE RMS MULTIMETER
87
LOLO
6
4
5
1
2
3
Produkspesifikasjoner
Merke: | Fluke |
Kategori: | Multimeter |
Modell: | 80K-15 |
Trenger du hjelp?
Hvis du trenger hjelp med Fluke 80K-15 still et spørsmål nedenfor, og andre brukere vil svare deg
Multimeter Fluke Manualer
30 August 2025
30 August 2025
30 August 2025
30 August 2025
30 August 2025
30 August 2025
30 August 2025
30 August 2025
30 August 2025
30 August 2025
Multimeter Manualer
- Multimeter Megger
- Multimeter Voltcraft
- Multimeter Klein Tools
- Multimeter PeakTech
- Multimeter EMOS
- Multimeter Velleman
- Multimeter Bearware
- Multimeter Tripp Lite
- Multimeter Noyafa
- Multimeter Rigol
- Multimeter Gossen Metrawatt
- Multimeter PCE Instruments
- Multimeter Owon
- Multimeter Extech
- Multimeter Abus
- Multimeter Uni-T
- Multimeter Flir
- Multimeter Vimar
- Multimeter Ideal
- Multimeter Alecto
- Multimeter Parkside
- Multimeter Metrix
- Multimeter Sanwa
- Multimeter Proline
- Multimeter Testboy
- Multimeter Kyoritsu
- Multimeter IWH
- Multimeter Trotec
- Multimeter Testec
- Multimeter Sauermann
- Multimeter Qian
Nyeste Multimeter Manualer
28 August 2025
28 August 2025
25 August 2025
25 August 2025
25 August 2025
24 August 2025
17 August 2025
16 August 2025
16 August 2025
16 August 2025