CO-220 English Instruction Sheet
Page
1
®
CO-220
Carbon Monoxide Meter
Instruction Sheet
Introduction
The CO-220 Carbon Monoxide Meter (hereafter referred to as
“the Meter”) detects the presence of carbon monoxide (CO) and
measures concentrations between 1-1000 parts per million
(PPM).
The Meter indicates the presence of carbon monoxide in two
ways:
• By a reading on the LCD in PPM.
• By a beeper tone.
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +3 402 67 200 1 - 5-
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at: www.fluke.com.
PN 690559 (English)
May 1998 Rev.2, 4/04
© 1998 F uke orporat on. r ghts reserved.-2004 l C i All i
All product names are trademarks of their respective companies.

CO-220 English Instruction Sheet
Page
2
W
Safety Information - Read First
⇒ Do not use the Meter as a personal safety monitor.
⇒ Learn and recognize the effects of CO poisoning.
0-1 PPM Normal background levels.
9 PPM ASHRAE Standard 62-1989 for living areas.
50 PPM OSHA enclosed space 8-hour average level.*
100 PPM OSHA exposure limit.*
200 PPM Mild headache, fatigue, nausea and dizziness.
800 PPM Dizziness, nausea and convulsions.
Death within 2 to 3 hours.
* U.S. Department of Labor, Occupational Safety & Health
Administration (OSHA) Regulation 1917.24:
The CO content in any
enclosed space shall be maintained at not more than 50 PPM
(0.005%). Remove employees from enclosed space if the CO
concentration exceeds 100 PPM (0.01 %).
What the Meter Does
The Meter indicates the presence of CO by a reading on the LCD
and a beeper tone.
Above 1000 PPM, the LCD reads “OL” to indicate an out of range
condition.
The beeper functions much like clicking of a Geiger counter:
• The higher the concentration of CO, the faster the beeper
sounds.
• Above 200 PPM, the beeper sounds continuously, while the
frequency of the tone increases with the concentration of CO.
Turning the Meter On
Install the battery. (See “Installing and Replacing the Battery.”)
To turn the Meter on, press the green button.
The Meter beeps twice, then performs a self-test. During the self-
test, the display counts down from “10” to “0”, while the Meter
emits a series of 2 beeps followed by a pause.
If the Meter passes the self-test, the Meter begins monitoring,
which is indicated by a beep every two seconds, and shows a
reading on the display.

CO-220 English Instruction Sheet
Page
3
Self-Test
The Meter tests the sensor and the battery each time it is turned
on and continuously monitors the battery. If the battery is low,
M is shown on the LCD.
If battery power is below the level required to power the Meter, or
if the sensor fails, the Meter emits a series of five beeps, followed
by a long pause, and another series of five beeps. The Meter
then turns itself off.
If after replacing the battery, the alarm condition persists, the
sensor may have failed. Contact an authorized Fluke service
center for information on how to replace the sensor.
Establishing a Baseline
Before testing for CO, establish a baseline against which the
Meter compares other environments.
Do this by turning on the Meter. If the Meter detects between
0-3 PPM of CO, the beeper sounds and the Meter establishes a
new baseline.
If the Meter detects more than 3 PPM, it does not establish a new
baseline, but starts taking CO measurements.
Storing Maximum Readings (Max/Hold)
In the Max/Hold mode, the Meter stores the maximum CO level it
has detected.
In this mode, the Meter continuously monitors the level of CO,
which is indicated by the beeper, while displaying only the
highest level detected on the LCD.
To enable or disable the Max/Hold mode, press the MAX button;
the beeper emits two quick tones when the Max/Hold mode is
enabled and disabled. When the Max/Hold mode is enabled,
“MAX” is shown on the LCD.
Auto-Back-light
If the ambient light is too low (making the LCD difficult to read),
the LCD back-light automatically turns on. The back-light
automatically turns off when sufficient light is present.
Automatic Power-O
The Meter automatically turns itself off after 20 minutes.
Produkspesifikasjoner
Merke: | Fluke |
Kategori: | Gassdetektor |
Modell: | CO-220 |
Trenger du hjelp?
Hvis du trenger hjelp med Fluke CO-220 still et spørsmål nedenfor, og andre brukere vil svare deg
Gassdetektor Fluke Manualer
30 August 2025
Gassdetektor Manualer
- Gassdetektor ORNO
- Gassdetektor Kidde
- Gassdetektor Joy-it
- Gassdetektor PCE Instruments
- Gassdetektor Vimar
- Gassdetektor AngelEye
- Gassdetektor Fantini Cosmi
- Gassdetektor Protector
- Gassdetektor Crowcon
- Gassdetektor Pentatech
- Gassdetektor Sauermann
Nyeste Gassdetektor Manualer
2 September 2025
31 August 2025
23 August 2025
22 August 2025
16 August 2025
12 August 2025
27 Mars 2025
29 Desember 2024
28 Desember 2024
22 Desember 2024