Operating Instructions
METRAHITIM XTRA BT
(M273A/D/W)
&
METRAHITIM E-DRIVE BT
(M274A/B)
&
METRAHITIM
TECH BT
(M272A/B)
Insulation Tester, Milliohmmeter,
TRMS Multimeter, Inter-Turn Short Circuit Tester
3-447-035-03
8/8.21

2 Gossen Metrawatt GmbH
Scope of Delivery (depending on instrument variant)
1 Multimeter with rubber holster
1HC40 hard case
(for multimeter and accessories)
(Z270K: black or Z270H: orange)
1 Quick-change, rechargeable lithium polymer battery with USB
power pack (5 V DC, 2 A) (Z270A or Z270G)
1 Probe (with start/stop and store/send function) (Z270S)
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
1 Cable set (1 pair of safety measurement cables, red/black,
with 4 mm test tips) (GTY362003P0002)
1 Pair of KC4 Kelvin clips (Z227A)
(METRAHIT IM XTRA BT and METRAHIT IM TECH BT only)
1 KC&S Kelvin clip and Kelvin probe (Z227C)
(METRAHIT IM E-DRIVE BT only)
1 DAkkS calibration certificate
1 Condensed operating instructions
* Comprehensive operating instructions available on the Internet
for download from www.gossenmetrawatt.com
1 IZYTRONIQ Business Starter
license (card with registration key
for software
Overview, Scope of Delivery
Key
X = standard
O = option
— = not possible, not intended
Overview of Included Features
1
Due to system design, the DC component displayed in the smallest measuring range
(300 mV) has an offset. Select the VDC function for precise measurement of the DC
component.
2
With optional temperature sensors
3
For 300,000 measured values, sampling rate adjustable from 0.1 seconds to 9 hours
4
30 kA = minimum breaking capacity
Accessories (sensors, plug inserts, adapters, consumable materials)
The accessories available for your instrument are checked for
compliance with currently valid safety regulations at regular inter-
vals, and are amended as required for new applications. Currently
up-to-date accessories which are suitable for your measuring
instrument are listed on our website along with photo, order num-
ber, description and, depending upon the scope of the respective
accessory, data sheet and operating instructions:
www.gossenmetrawatt.de
Accessories Type Article
No.
M273S M274S M272S
METRAHIT IM XTRA BT
M273D X
METRAHIT IM E-DRIVE BT M274B X
METRAHIT IM TECH BT M272B X
Quick-change, rechargeable
lithium polymer battery with
USB power pack M27x
Z270A/
Z270G X X X
USB charger with 4 inter-
changeable primary connec-
tors (for Z270A/ Z270G) M27x Z270 L O O O
Probe with keys Z270S Z270S X X —
Cable set
KS17-2
GTY3620
03P0002 X X X
1 pair of Kelvin clips KC4 Z227A X O X
1 pair of Kelvin probes KC27 Z227B O O O
1 Kelvin clip &
1 Kelvin probe KC&S Z227C O X O
Concentric Kelvin probes for
4-wire measurement KCC Z227O O O O
Cable reel for 4-wire measure-
ment, 100 meters KCV100 Z227E O O O
Hard case
Black
Orange HC40
Z270K
Z270H
X
X
X
Magnetic holder and
Velcro fastener
HIT-Clip Z117A O O O
COIL adapter, 10 μH … 50
mH
COIL TEST
ADAPTER Z270F O O —
COIL adapter, 10 μH … 5 H
COIL ADAPTER
XTRA Z270M O O —
Set of test probes with
alligator clips for
COIL Adapter XTRA
KSC-3L Z110C O O —
Adapter cable, 4 mm plug
to 6 mm socket AK-4M/6F Z110L O O O
Functions expansion to 16 test
sequences with up to 63 test
steps each
Sequence
functions
Expert
Z270P
OOO
IZYTRONIQ Business Starter li-
cense S101S & Z956A
S101S &
Z956A X X X
Function METRAHIT IM
XTRA BT
E-DRIVE BT
METRAHIT
IM TECH BT
V
DC
(Ri 9 M)
V
AC
/ Hz TRMS (Ri 9 M)
filter
filter
V
AC+DC
TRMS (Ri 9 M)
1
filter
filter
V
AC+DC
TRMS (Ri 1 M)
R
ISO
range (interference voltage)
Hz (V
AC
)
... 300 kHz
... 300 kHz
V
AC, AC+DC
bandwidth
100 kHz
100 kHz
A
DC, AC, AC+DC
/ Hz TRMS
10 nA ... 1 A
10 nA ... 1 A
Fuse F1, current measuring function
1 A/1000 V - 30
kA
4
1 A/1000 V -
30 kA
4
Current sensor transformation ratio
1 mV : 1
• 10 •
100 • 1000 mA
1 mV : 1 • 10 •
100 • 1000 mA
Hz (A AC)
... 30 kHz ... 30 kHz
Insulation resistance RISO: Test Voltages
50
• 100 • 250 •
500 • 1000 V
Inter-turn short circuit test (1 kV) with COIL adapter
Option
Duty cycle measurement as %
RPM measurement
Resistance Rlo with 200 mA per EN 61557
Milliohm with 4-wire method, m with 200 mA
Milliohm with 4-wire method, m with 1 A
pulse
Fuse F2, Rlo measuring function
315 mA/1000 V –
30 kA
4
Resistance
Continuity
Diode ... 4.5 V
Temperature: °C/°F TC type K and Pt100/1000
2
Capacitance
Min-Max / data hold
Test sequence
1
(with 10 steps)
Expert sequence functions
Option Option
64 MBit memory
3
Bluetooth
®
interface
3.5" TFT color graphic display
Probe with start/stop and send/store keys
Quick-change battery with USB charging
Protection
IP 52
IP 52
Measuring category
1000 V CAT III,
600 V CAT IV
1000 V CAT III,
600 V CAT IV
1 kHz
1 kHz
1 kHz
1 kHz

Gossen Metrawatt GmbH 3
Contents Page Contents Page
1 Safety Instructions ............................................................5
2 Applications.......................................................................6
2.1 Intended Use / Use for Intended Purpose .................................... 6
2.2 Use for Other than Intended Purpose........................................... 6
2.3 Liability and Guarantee................................................................ 6
2.4 Opening the Instrument / Repairs................................................ 6
2.5 Manufacturer’s Guarantee........................................................... 6
3 Documentation ..................................................................6
3.1 Product Variants .......................................................................... 6
3.2 Firmware Version......................................................................... 6
3.3 Symbols ....................................................................................... 6
4 Operating Overview ...........................................................7
4.1 Connections, Keys, Rotary Switch, Symbols
(depending on instrument variant) .............................................. 7
4.2 Symbols Used in the Digital Display ............................................ 8
4.3 Rotary Switch Position Symbols
(depending on instrument variant) ............................................. 8
4.4 Symbols on the Instrument.......................................................... 9
5 Initial Startup...................................................................10
5.1 Battery Pack .............................................................................. 10
5.2 Switching the Instrument On ..................................................... 11
6 System Settings .............................................................. 12
6.1 Digital Keyboard ........................................................................ 12
6.2 Firmware ................................................................................... 12
6.3 Language Selection ................................................................... 12
6.4 Time and Date Settings ............................................................. 12
6.5 Digital Display Settings.............................................................. 13
6.5.1 Brightness ................................................................................... 13
6.5.2 Display Profile.............................................................................. 13
6.6 Set Display of Leading Zeros ..................................................... 13
6.7 Configuring Password Protection .............................................. 13
6.7.1 Selecting or Changing a Password ................................................ 13
6.7.2 Deactivating Password Protection.................................................. 13
6.8 Viewing the Instrument’s Name................................................. 14
6.9 Installing a Functions Expansion ............................................... 14
6.10 Switching the Instrument Off..................................................... 14
6.10.1 Switching the Instrument Off Manually........................................... 14
6.10.2 Automatic Shutdown (auto-off)...................................................... 14
6.11 Default Settings (instrument reset)............................................ 14
7 Control Functions ............................................................16
7.1 Help............................................................................................ 16
7.2 Selecting Measuring Functions and Measuring Ranges............ 17
7.2.1 Automatic Range Selection ........................................................... 17
7.2.2 Manual Range Selection ............................................................... 17
7.2.3 Quick Measurements ................................................................... 17
7.3 Zero Offset / Relative Measurements ........................................ 17
7.4 Display (TFT) .............................................................................. 18
7.4.1 Digital Display .............................................................................. 18
7.4.2 Analog Display ............................................................................. 18
7.5 Measured Value Storage – Data Function
(Auto-hold/Compare) ................................................................. 18
7.5.1 Saving Minimum and Maximum Values – “MinMax”....................... 19
7.6 Measured Value Memory – STORE Function. ............................ 20
7.6.1 Remote read-out and storage via the Z270S probe
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only) .......... 20
7.6.2 Remote Read-Out and Storage via the PC – PUSH/PRINT Function.. 20
7.7 Measurement Data Recording ................................................... 21
7.7.1 One-Time Only (manual) Storage................................................... 21
7.7.2 Measurement Series (automatic storage)....................................... 21
7.7.3 Memory....................................................................................... 23
8 Measurements.................................................................24
8.1 Enabling Parameter Changes .................................................... 24
8.2 Insulation Resistance Measurement – RISO Function
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only) .... 24
8.2.1 Preparing for Measurement .......................................................... 24
8.2.2 Performing Insulation Measurement.............................................. 26
8.2.3 Ending the Measurement and Discharging..................................... 26
8.3 Inter-Turn Short Circuit Measurement – COIL Function
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only) ......... 26
8.3.1
Inter-Turn Short Circuit Measurement with the COIL TEST ADAPTER
.26
8.3.2 Inter-Turn Short Circuit Measurement with the ADAPTER XTRA ...... 30
8.4 Absorption Index Measurement – DAR
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only) .... 34
8.5 Polarization Index Measurement – PI
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only) .... 35
8.6 Voltage Measurement................................................................ 36
8.6.1 Alternating Voltage and Frequency Measurement V AC and Hz with
Selectable Low-Pass Filter............................................................ 36
8.6.2 Duty Cycle Measurement – Duty AC
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only) .......... 37
8.6.3 RPM Measurement – RPM AC
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only) .......... 37
8.6.4 Direct and Pulsating Voltage Measurement, V DC and V (AC+DC)... 39
8.7 Resistance Measurement “” .................................................. 40
8.8 Capacitance Measurement F..................................................... 41
8.9 Temperature Measurement with Resistance Thermometers –
Temp RTD .................................................................................. 42
8.10 Temperature Measurement with Thermocouple – Temp TC ..... 43
8.11 Continuity Test .......................................................................... 44
8.12 Diode Testing with Constant Current of 1 mA .......................... 44
8.13 Milliohm Measurement – Rlo (2-wire measurement)
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only) ......... 45
8.14 Milliohm Measurement – m/4 (4-wire measurement) ........... 46
8.14.1 Compensation of Cable Resistance ............................................... 46
8.14.2 Thermovoltage Compensation....................................................... 47
8.14.3 Milliohm Measurement with 200 mA or 20 mA DC [m
8.14.4 Milliohm Measurement with 1 A Pulsating Measuring Current
(automatic thermovoltage correction at 3 … 300 m)................... 48
8.15 Current Measurement................................................................ 49
8.15.1 Direct and Pulsating Current Measurement, Direct Connection, –
A DC and A (AC+DC) ................................................................... 49
8.15.2 Alternating Current and Frequency Measurement,
Direct Connection, – AAC and Hz.................................................. 50
8.15.3 Direct and Pulsating Current Measurement with
Current Clamp Sensor – ADC und A (AC+DC)................................ 51
8.15.4 Alternating Current Measurement with Current Clamp
Sensor – AAC and Hz................................................................... 52
8.16 Measuring Sequences – Test Sequences.................................. 53
9 Interface Operation and Software ...................................56
9.1 Bluetooth®................................................................................ 56
9.2 Software for Receiving and Evaluating Data ............................. 57
9.2.1 METRAHIT IM Data Reader (PC) ................................................. 57
9.2.2 Terminal Program (PC) ................................................................. 57
9.2.3 METRALOG App (smartphone and tablet)....................................... 58
9.2.4 Sequence Manager: Software for Test Sequences ....................... 58
10 Characteristic Values.......................................................61
11 Maintenance and Calibration...........................................65
11.1 Displays – Error Messages ........................................................ 65
11.2 Fuses ......................................................................................... 65
11.3 Housing Maintenance................................................................ 65
11.4 Measurement Cables................................................................. 65
11.5 Recalibration ............................................................................. 65
Produkspesifikasjoner
Merke: | Gossen Metrawatt |
Kategori: | Multimeter |
Modell: | METRAHIT IM E-DRIVE BT |
Trenger du hjelp?
Hvis du trenger hjelp med Gossen Metrawatt METRAHIT IM E-DRIVE BT still et spørsmål nedenfor, og andre brukere vil svare deg
Multimeter Gossen Metrawatt Manualer
2 August 2025
2 August 2025
2 August 2025
2 August 2025
Multimeter Manualer
- Multimeter Fluke
- Multimeter Megger
- Multimeter Voltcraft
- Multimeter Klein Tools
- Multimeter PeakTech
- Multimeter EMOS
- Multimeter Velleman
- Multimeter Bearware
- Multimeter Noyafa
- Multimeter Rigol
- Multimeter PCE Instruments
- Multimeter Owon
- Multimeter Extech
- Multimeter Abus
- Multimeter Uni-T
- Multimeter Vimar
- Multimeter Ideal
- Multimeter Metrix
- Multimeter Sanwa
- Multimeter Proline
- Multimeter Testboy
- Multimeter Kyoritsu
- Multimeter IWH
- Multimeter Testec
- Multimeter Qian
Nyeste Multimeter Manualer
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
29 Mars 2025
26 Mars 2025
14 Mars 2025