Gossen Metrawatt METRATESTER 5+ Bruksanvisning

Les nedenfor 📖 manual på norsk for Gossen Metrawatt METRATESTER 5+ (28 sider) i kategorien måleinstrument. Denne guiden var nyttig for 6 personer og ble vurdert med 4.5 stjerner i gjennomsnitt av 2 brukere

Side 1/28
Operating Instructions
3-348-580-15
28/9.16
METRATESTER 5+
Tester for DIN VDE 0701-0702
2 GMC-I Messtechnik GmbH
(1) Mains plug
(take-up spool at back of housing
for mains cable)
(2) Signal lamp PE for testing the
mains protective conductor
(3) Alligator clips for attachment
to the test probe (3a)
(3a) Test probe
(4) Contacting surface for contact finger
(5) Measuring function selector switch
R
SL
Protective Conductor Resistance
R
ISO
Insulation Resistance
I
EA
Equivalent Leakage Current
I
A
Contact or Leakage Current
(for confirmation of absence of voltage)
I
Diff
Residual Current
I
Netz
Load current at mains outlet
U
Netz
Mains Voltage
(6) Connector jack/terminal for testing for
contact current at conductive parts at
the DUT
(7) Mains outlet
(8) Test socket
(9) Connector jack/terminal for DUT
phase conductors (wired parallel
to the socket)
(10) Connector jack/terminal for DUT
protective conductor (wired parallel
to test socket)
(11) LCD display (description see page 24)
(12) Carrying handle
(13) Error lamp
(8) (9) (10) (11) (12)
(7)
(6)
(5)
(4)
(3)
(2)
(1)
(13)
(3a)
GMC-I Messtechnik GmbH 3
Display
Display text subject to change without notice.
Meanings of Symbols on the Instrument
Continuous, doubled or reinforced
insulation
Warning concerning a point of danger
(attention: observe documentation!)
Indicates EC conformity
This device may not be disposed with the trash. For further details on the WEEE
marking, please refer to our website www.gossenmetrawatt.com and enter
search key ’WEEE’.
I
Diff
mA I
Netz
A
MΩ
R
SL
>0.3
>1.0
R
ISO
<0.5
<2.0
I
EA
>7.0
I
A
>0.25
>0.5
mA
mA
Ω
I
Diff
mA I
Netz
AU
Netz
V
Insulation
Resistance
PE
Resistance
Equivalent
Leakage Current
Leakage Current/
Contact Current
Residual
Current
Secondary
Current
Indication of Limit Values
Mains
Voltage
KS 13 Accessory Cable Set
!