Tripp Lite HCRK-1 Bruksanvisning
Tripp Lite
batteri
HCRK-1
Les nedenfor 📖 manual på norsk for Tripp Lite HCRK-1 (16 sider) i kategorien batteri. Denne guiden var nyttig for 2 personer og ble vurdert med 4.5 stjerner i gjennomsnitt av 2 brukere
Side 1/16

1
Owner’s Manual
Medical-Grade
Mobile Power
Retrofit Kit
Model: HCRK
Kit Includes: Power Supply Module, Battery Module, Remote User Interface (RUI),
PowerAlert Data-Saving Software, Mounting Hardware and Cables.
Fully tested, certified and compliant with:
• UL 60601-1 • CAN/CSA-C22.2 No. 601.1-M90 • UL 1778 by
1111 W. 35th Street, Chicago, IL 60609 USA
www.tripplite.com/support
Copyright © 2010 Tripp Lite. All rights reserved.
Important Safety Instructions 2
Mounting 3
Connection / Start-Up 7
Operation 9
Features 10
Troubleshooting 12
Storage & Service 13
Safety & Regulatory Statements 14
Warranty & Warranty Registration 14
WARRANTY
REGISTRATION:
register online today for a
chance to win a FREE Tripp Lite
product—www.tripplite.com/warranty
200910240 93-2650-EN.indd 1 6/16/2010 5:06:26 PM

2
Important Safety Instructions
Tripp Lite’s Medical-Grade, Mobile Power Retrofit Kit includes two modules: a Power Supply Module and
a Battery Module. When the two modules are connected (and attached to a remote user interface), they
provide AC power for a variety of equipment on mobile carts. The following instructions refer to the kit as
a single unit as well as its component parts.
SAVE THESE INSTRUCTIONS
This manual contains instructions and warnings that should be followed during the installation, operation
and storage of all Tripp Lite Medical-Grade, Mobile Power Retrofit Kits. Failure to heed these warnings
may affect your warranty.
Power Supply Module & Battery Module Location Warnings
• Your Power Supply Module incorporates fusing only in the ungrounded phase conductor. UL 60601-1
certified Power Supply Modules must not be used in countries other than the United States and Canada
and must be used only in health care facilities on grounded systems where conditions of maintenance
and supervision ensure that only qualified persons will service the electrical distribution system.
• Install your Power Supply Module and Battery Module indoors, away from excess moisture or heat, dust
or direct sunlight and conductive contaminants.
• Although the Power Supply Module is designed to support computer equipment, do not install it in direct
contact with magnetic storage media, as this may result in data corruption.
• For best performance, keep the indoor temperature between 32º F and 104º F (0º C and 40º C).
• Leave adequate space around all sides of the Power Supply Module and Battery Module for
proper ventilation.
Power Supply Module Connection Warnings
• Do not operate the Power Supply Module without connecting it to a Battery Module and the Remote
User Interface (RUI). The Power Supply Module will not operate (with or without utility power) until
the Battery Module and RUI are connected.
• Do not allow objects to contact the Power Supply Module’s two DC input terminals. Do not short or
bridge these terminals together. Serious personal injury or property damage could result.
• Connect your Power Supply Module directly to a properly grounded AC power outlet. Do not plug the
Power Supply Module into itself; this will damage the unit. The Power Supply Module features a
hospital-grade plug. If the plug is connected to a wall outlet that is not hospital-grade, the unit may not
be properly grounded.
• Do not modify the Power Supply Module’s plug, and do not use an adapter that would eliminate
the Power Supply Module’s ground connection.
• Do not use extension cords to connect the Power Supply Module to an AC outlet.
• If the Power Supply Module receives power from a motor-powered AC generator, the generator must
provide clean, filtered, computer-grade output.
Equipment Connection Warnings
• Use of this equipment in life support applications where failure of this equipment can reasonably be
expected to cause the failure of the life support equipment or to significantly affect its safety or
effectiveness is not recommended. Do not use this equipment in the presence of a flammable anesthetic
mixture with air, oxygen or nitrous oxide.
• The Power Supply Module is designed to accept connection of equipment mounted on the same mobile
cart. Connecting equipment which is not mounted on the same cart as the Power Supply Module could
pose a safety hazard.
• Do not connect surge suppressors or extension cords to the output of your Power Supply Module. This
may damage the Power Supply Module and may affect the surge suppressor and Power Supply Module
warranties.
200910240 93-2650-EN.indd 2 6/16/2010 5:06:26 PM

3
Battery Module Warnings
• Do not allow objects to contact the Battery Module’s two DC input terminals. Do not short or bridge
these terminals together. Serious personal injury or property damage could result.
• It is normal for sparks to occur when connecting the Battery Module to the Power Supply Module.
• Due to the possibility of dangerous arcing, do not unplug the Battery Module from the Power Supply
Module while the Power Supply Module is operating from battery power (whenever the cart is mobile
and the Power Supply Module is not plugged into a live utility wall outlet, or during a power outage).
• Your Battery Module does not require routine maintenance. Do not open your Battery Module for any
reason. There are no user-serviceable parts inside. The batteries in the Battery Module will eventually
wear out and be unable to provide adequate run times. Internal battery replacement must be
performed by factory authorized service personnel. Because batteries present a risk of electrical shock
and burn from high short-circuit current, service personnel should observe proper precautions. Unplug
the Battery Module before performing battery replacement. Do not short or bridge battery terminals with
any object. Use tools with insulated handles. Remove metal objects such as watches and rings before
working with batteries. Do not open batteries. Replace batteries only with sealed lead-acid batteries of
the same number, type, voltage and amp-hour capacity. Contact Tripp Lite for details.
• Tripp Lite offers additional replacement Battery Modules (model: BP12V82). Contact Tripp Lite
for details.
• The Battery Module’s internal batteries are recyclable. Refer to local codes for disposal requirements, or
visit www.tripplite.com/UPSbatteryrecycling for recycling information. Do not dispose of the batteries in a
fire.
• If connecting a replacement Battery Module to the Power Supply Module, the Battery Module must have
the same voltage as the Power Supply Module. Use only Tripp Lite Battery Modules
(model: BP12V82) with the Power Supply Module.
Labeling Symbols
AC Voltage: V~
DC Voltage: V
Neutral Fusing:
Electric Shock Hazard:
Important Safety Instructions continued
Mounting
WARNING!
Mount the Power Supply Module and Battery Module BEFORE connection.
The mounting brackets are only designed to support a Tripp Lite Power Supply Module and a
single Battery Module. Do not mount or place any other devices on the mounting brackets.
The mounting brackets will not support the additional weight and may fail.
Failure to follow these warnings may lead to product damage and/or risk of personal bodily harm.
The included mounting materials are designed to adapt the Mobile Power Retrofit Kit to fit multiple cart
configurations. See diagrams on the following pages to locate the mounting procedure that matches your
specific cart configuration. The mounting procedures include:
• Mounting on a Center Pole
• Mounting on a Bottom Shelf (Replacing Existing Shelf)
• Mounting on a Bottom Shelf (Using Existing Shelf)
The user must determine the fitness of the mounting materials, hardware and procedures before mounting.
If the materials, hardware or procedures are not suitable for your application, contact Tripp Lite for
assistance.
200910240 93-2650-EN.indd 3 6/16/2010 5:06:27 PM
Produkspesifikasjoner
Merke: | Tripp Lite |
Kategori: | batteri |
Modell: | HCRK-1 |
Trenger du hjelp?
Hvis du trenger hjelp med Tripp Lite HCRK-1 still et spørsmål nedenfor, og andre brukere vil svare deg
batteri Tripp Lite Manualer
17 August 2025
17 August 2025
17 August 2025
17 August 2025
17 August 2025
17 August 2025
17 August 2025
17 August 2025
17 August 2025
17 August 2025
batteri Manualer
Nyeste batteri Manualer
16 August 2025
6 August 2025
5 August 2025
4 August 2025
4 August 2025
3 August 2025